True regarding anaesthesia apparatus –
Correct Answer: An oxygen failure safety device operate even in the absences of nitrous oxide
Description: Oxygen failure safety device operates to detect oxygen supply pressure (not N2O pressure). So, if O2 supply pressure is low it operates to interrupt the flow of N2O.